Ivoclar Programat P95 Manual [repack]
To safely unpack and establish your workspace, follow these sequential steps from the Ivoclar Download Center guidelines : IVOCLAR VIVADENT Programat P95 buy on Bimedis
| Error Code | Meaning | Operator Action | | :--- | :--- | :--- | | | Thermocouple broken (infinite resistance) | Replace type B TC. | | E-23 | Heating rate too slow (element aging) | Run regeneration cycle (1500°C, 2h, no load). | | E-45 | Vacuum leak | Check door seal and hose connections. | | E-67 | Cooling fan stalled | Clean fan blades; check 12V DC supply. | | E-88 | Overtemperature (>1650°C) | Immediate reset. Check SSR relay for short circuit. | Ivoclar Programat P95 Manual
Never delete factory program 0 – it contains the service and calibration menu. To safely unpack and establish your workspace, follow
The core utility of the Ivoclar Programat P95 manual lies in its explanation of the user interface and programming capabilities. The Programat P95 is designed to handle a variety of materials, including Ivoclar’s renowned IPS e.max and classic metal-ceramics. The manual serves as a comprehensive guide to navigating the control panel, detailing how to select pre-set programs and how to customize firing cycles. For a dental technician, understanding the parameters of , heating rate , final temperature , and vacuum levels is essential. The manual decodes these variables, allowing the user to tailor the firing cycle to the specific thermal expansion coefficients of the ceramic material being used. Without the manual’s detailed charts and programming logic, a technician might inadvertently misuse the thermal cycle, leading to porous or cracked restorations. | | E-67 | Cooling fan stalled |
The furnace failed to reach the required vacuum level. Check the lift O-ring seal, vacuum hose connections, or pump oil.
Maximum firing temperature of 1200°C (2192°F).
The Programat P95 is designed for the standard firing of dental ceramic materials, including classic metal-ceramics and certain low-fusing glass-ceramics. Specification 110–120 V or 220–240 V (Model dependent) Maximum Firing Temp 1200°C (2192°F) Display Digital LED / Segment Display Muffle Technology High-efficiency heating element with quartz tube protect Vacuum Control Integrated electronic vacuum control 🚀 Installation and First-Time Setup